Description Ultra-Low Differential Pressure Transmitter | PMM06-D Measurement Measuring element Thermal mass type Input Differential Measuring range ±10, ±25, ±50, ±100 Pa Output Output 4 … 20 mA / 0 … 10 V / Relay / RS-485 Signal connection M12, 3-wire Warm-up time 60 sec Response time t90 ≦ 6 sec Load resistance Current output: ≦ 500 Ω Voltage output: ≧10 KΩ Communication Communication methods & protocol RS-485 Modbus RTU RS-485 baud rate 9600, 19200, 38400, 57600, 115200 bps Accuracy Accuracy ±(1% F.S. 0.1 Pa) Temperature influence ±0.5% F.S./°C Environmental Measuring medium Air Operating Temp. & Humid. 0 … 50°C / 20 …90%RH(Non-condensing) Storage Temp. -25 … 60°C Operating pressure 16 bar Electrical Power supply DC 24 V ±10% Current consumption 24 V : 60 mA Relay capacity Max current: 6 A Max voltage: DC 24 V (DC 36 V Max) Electrical connection M12 8 PIN connector Installation Installation Compatible with Ø8 PVC PTFE pneumatic tubing Display Display readout Diff. pressure value with two decimal places Decimal point Button Sampling time 1 cycle/sec Diff. pressure unit Pa, mbar, hPa, mmWS, inH2O, mmHg Response time adjustment range 0.5 … 300 sec Certification Certification CE Protection IP rating IP65(Housing) Electrical protection ■ Over-voltage ■ Reverse polarity Material Housing Aluminium alloy / Plastic Weight 207 g Ultra-Low Differential Pressure Transmitter | PMM06-D